NR AKMO

AU Schell,J.; Wulfert,M.; Riesner,D.

TI Detection of point mutations by capillary electrophoresis with temporal temperature gradients

QU Electrophoresis 1999 Oct; 20(14): 2864-9

PT journal article

AB By combining the advantages of capillary electrophoresis and temperature gradient gel electrophoresis, a method was developed to detect point mutations in polymerase chain reaction (PCR) fragments. Increasing and decreasing temporal temperature gradients were established by means of a computer-controlled Peltier module. Native and denaturing conditions were achieved by cooling to 25 degrees C and heating to 70 degrees C, respectively, a thermostating liquid surrounding the capillary. To separate nucleic acid fragments, a sieving media, containing 4% linear polyacrylamide, 1 x Tris borate EDTA buffer (TBE) and 6 M urea, was found appropriate. Renewal of the sieving matrix before each run significantly improved the reproducibility of fragment separation. The ability of this capillary electrophoresis system to detect point mutations is demonstrated with the human prion-protein gene.

MH Animal; DNA/analysis/*genetics; Electrophoresis, Capillary/methods; Human; *Point Mutation; Temperature

AD Institut für Physikalische Biologie and Biologisch-Medizinisches Forschungszentrum, Heinrich-Heine-Universität, Düsseldorf, Germany.

SP englisch

PO Deutschland

EA pdf-Datei

Autorenindex - authors index
Startseite - home page